TECHNICAL SPECIFICATION Interferential: Continous Sine Wave Muscle stimulator / Russian: Continous Premodulated Sine wave Galvanic; DC & Monophasic
WAVEFORM TENS: Symmetric Bi-Phasic Square Wave. IFT: CH1/CH3 = 4000Hz, CH2/CH4 = Selectable 1- 250Hz Russian: CH1/Ch3 = 2500Hz, CH2/CH4 = 2550Hz (Fix) Interrupted Galvanic: 1Hz (Fixed)
FREQUENCY TENS: Selectable 1-150Hz IFT: 125μS, Russian 200 μS TENS: Selectable 50-400 μS
IMPULSE Interrupted Galvanic: 2ms Width Output Vector Rotation Scanning IFT: 100% Scan Contraction Time: Selectable 1-30seconds Relaxation Time: Selectable 0-30seconds Number of Electrodes: 8 Electrodes/ 2 leadwires
TOLERANCES Pulse Rate (Hz): + 10%, Pulse Width (μS): + 10%, Output: + 10%
CLINICAL INTERFERENTIAL STIMULATOR BLD AP439
OUTPUT INTENSITY IFT: 60Vpp @ Resistive Load of 500 ohms Russian: 60Vpp @ Resistive Load of 500 ohms Galvanic: 16Vpp @ Resistive Load of 500 ohms Interrupted Galvanic: 24Vpp @ Resistive Load of 500ohms TENS: 70Vpp @ Resistive Load of 500 ohms
MODES 8 (Interferential, Premod Bi-polar, Russian, Galvanic, Interrupted Galvanic, TENS-Conventional, TENS-Burst and TENS-Modulation) Amplitude (mAmp) 120 mA @ resistive load of 500 ohms. Timer: Selectable 1-60mins Digital: Yes
© 2010 Source Medical Devices